Diamond Type

We design and manufacture our Diamond Type Gas Outlets (Terminal Units) to last and to ensure a constant and reliable medical gas flow for the patient and the medical equipment.

All outlets are 100 % tested for flow, and leakage and are cleaned for oxygen services before shipping.

Our Diamond Type Gas Outlets, also known as Sil-Ohio gas outlets, are designed by industrial designers to create a pleasant and appealing environment for both patients and medical staff.

Downloads:

  • Accepts Silbermann diamond-
    type gas-specific
     adapters.
  • All outlets are 100% hydrostatic
    tested 
    for leaks. The outlets are factory-
    assembled, degreased, and cleaned for
    medical gas service.
  • Indexed to eliminate interchangeability
    of gas services.
  • 360 deg swivel inlet pipe for ease of
    installation.
  • Low maintenance due to use of VITON
    seals in front assembly (latch value).


Medical Gas System Standards:

  • NFPA99, EN ISO 7396-1 requirements.

Applicable Standards:

  • NFPA99,  ISO 9170-1.

Silbermann’s Quality Management System:

  • CE marked by the notified body TÜV-SÜD.


  • The medical gas outlet shall be
    SILBERMANN Diamond type
    manufactured according to ISO 9170-1.
  • The outlet shall be gas-specific by
    indexing each front service panel
    to a gas-specific pin indexed

    rough-in module.
  • The outlet of positive and negative
    pressure shall consist of a primary
    and secondary check
     valves to allow
    testing and on-line 
    maintenance, i.e.,
    the front service panel shall
     be
    removable for service without
    the need to 
    shut down the gas supply.
  • Front latch valve assembly shall
    incorporate an anodized aluminum front panel.
  • Outlet shall be manufactured with
    175 mm long, type K, 12.7 mm (1/2”) OD
    copper pipe.

    Rough-in pipe shall be silver brazed to
    the outlet body.
  • With the rough-in mounted, the outlet
    shall be
     depth adjustable to compensate
    for various wall thicknesses.
  • The quick connect outlet
    shall comply with ISO 9170-1
    standard type and accept only

    corresponding Silbermann Diamond Gas
    Specific Adapter.
  • Pressure drop of less than 20 Kpa
    through 
    the outlet at 120 LPM and
    350 Kpa (3.5 Bar) 
    inlet pressure.
  • When fully open, the outlet allows a flow
    of up 
    to 600LPM and pressure
    of 400 KPa (4 Bars).