Diamond Type

We design and manufacture our Medical Gas Outlets (Terminal Units) to last and to ensure a constant and reliable medical gas flow for the patient and the medical equipment.

All outlets are 100 % tested for flow, leakage and are cleaned for oxygen services prior to shipping.

Our outlets are designed by industrial designers to ensure an appealing and pleasant environment for the patient and the medical staff.

Downloads:

   Accepts Silbermann diamond-
      type gas-specific
 adapters.

   All outlets are 100% hydrostatic
     tested 
for leaks. The outlets are factory
     assembled, degreased, and cleaned for
     medical gas service.

   Indexed to eliminate interchangeability of
     gas services.

  360 deg swivel inlet pipe for ease of
     installation.

   Low maintenance due to use of VITON
     seals in front assembly (latch value).


Medical Gas System Standards:

      NFPA99, EN ISO 7396-1 requirements.

Applicable Standards:

    NFPA99,  ISO 9170-1.

Silbermann’s Quality Management System:

    CE marked by the notiied body TÜV-SÜD.


   The medical gas outlet shall be
     SILBERMANN Diamond type manufactured
     according to ISO 9170-1.

   The outlet shall be gas-specific by indexing
     each front service panel to a gas-specific pin
     indexed rough-in module.

   The outlet of positive and negative pressure
     shall consist of a primary and secondary
     check valves to allow testing and on-line
     maintenance, i.e., the front service panel
     shall
 be removable for service without
     the need to 
shut down the gas supply.

   Front latch valve assembly shall incorporate
     an anodized aluminum front panel.

   Outlet shall be manufactured with 175 mm
      long, type K, 12.7 mm (1/2”) OD
      copper pipe.

     Rough-in pipe shall be silver brazed to the
     outlet body.

   With the rough-in mounted, the outlet shall
     be
 depth adjustable to compensate for
     various wall thicknesses.

   The quick connect outlet shall comply with
     ISO 9170-1 standard type and accept only
     corresponding Silbermann Diamond Gas
     Specific Adapter.

   Pressure drop of less than 20 Kpa through
     the outlet at 120 LPM and 350 Kpa (3.5 Bar)
     inlet pressure.

   When fully open, the outlet allows a flow
     of up 
to 600LPM and pressure of 400 KPa
     (4 Bars).