ENV 737-6 Type
We design and manufacture our Medical Gas Outlets (Terminal Units) to last and to ensure a constant and reliable medical gas flow for the patient and the medical equipment.
All outlets are 100 % tested for flow, leakage and are cleaned for oxygen services prior to shipping.
Our outlets are designed by industrial designers to ensure an appealing and pleasant environment for the patient and the medical staff.
Downloads:
■ Accepts ENV 737-6 gas-specific adapters.
■ All medical gas outlets are 100%
hydrostatic tested for leaks.
The outlets are factory-assembled,
degreased, and cleaned for
medical gas service.
■ Indexed to eliminate interchangeability
of gas services.
■ 360-degree swivel inlet pipe for
ease of installation.
■ Double position allowing parking of
male adapter in the outlet.
■ Low maintenance due to use of VITON
seals in front assembly (latch value).
Medical Gas System Standards:
■ NFPA99 & EN ISO 7396-1 requirements.
Applicable Standards:
■ ISO 9170-1, ENV 737-6.
Silbermann's Quality Management System:
■ CE marked by the notified body TÜV-SÜD.
■ The medical gas outlet shall be
SILBERMANN ENV type manufactured
according to ISO 9170-1 ENV 737-6.
■ The outlet shall be gas-specific
by indexing each front service panel
to a gas-specific pin-indexed rough-in
module.
■ The outlet of positive and negative
pressure shall consist of a primary
and secondary check valve to allow
testing and online maintenance.
■ Outlet shall be manufactured with
175 mm long, type K, 12.7 mm (1/2”)
OD copper pipe. Rough-in pipe shall be
silver brazed to the outlet body.
■ With the rough-in mounted, the outlet
shall be depth adjustable to compensate
for various wall thicknesses.
■ The quick-connect outlet shall comply
with ISO 9170-1 standard type and accept
only corresponding ENV 737-6 Gas
Specific Adapter.
■ Pressure drop of less than 20 Kpa
through the outlet at 120 LPM and 350 Kpa
(3.5 Bar) inlet pressure.
■ When fully open, the outlet allows a low
of up to 600LPM and pressure of
400 KPa (4 Bars).