Gas Scavenging Inlet (AGSS)

The Silbermann jet scavenging inlet is designed for safe removal of surplus anesthetic gas and explosive gas mixtures from the operating theater.

The Silbermann jet scavenging inlet is designed to meet the latest and most rigid NFPA99 requirements .The scavenging inlet assembly uses compressed air which creates suction (Venturi) and propels the contaminated air to the outside atmosphere.

The jet scavenging inlet can be supplied recessed, panel mounted or exposed ,with or without hose assembly

Downloads:

■  Operation based on compressed air -
     No need for a central AGSS system

■  When the probe is connected, green
    pneumatic indicator showing the system
    is in use

■  Very quiet operation

■  Automatic shut off when the probe is
    disconnected

■  Flow and vacuum adjustment

■  The locking mechanism quick connect

 

Medical Gas System Standards:

■  NFPA99
■  ISO 7396
■  G-01
■  HTM 02-01

 

Silbermann’s Quality Management System:

■  EN ISO 13485-2016
■  ISO 9001-2015
■  CE marked

■  Inlet pressure: 3.5-5 bars

■  Inlet flow: 40 LPM

■  Free air suction capacity:
    Adjustable 35 -80 LPM

■  Vacuum level: Adjustable 20-50 cmH2O

■  Inlet pipe diameter: OD 3/8”

■  Exhaust pipe diameter: OD 7/8”

■  Material of construction:
    ■  Box: Baked painted Steel
    ■  Front Panel: Anodized Aluminum
    ■  Injector: Brass