Area Alarm Panel

The Silbermann Advanced Area Alarm
System displays digitally and graphically
on a colorful 12” touch screen the
medical gas information from the
manifold systems. 

This Xanadu touchscreen model uses
Hi security technology that protects
The information displayed and
transmitted, from malicious attacks.

The Xanadu Area Alarm System complies
with the NFPA99, EN ISO 7396-1, and HTM02-01.

 

Xanadu Touch screen Area and main alarm panel video.

Downloads:

  • LCD 12” touch screen display.
  • Display of 1-10 gas alarms on screen.
  • Aesthetics & smooth design without
    buttons for infection control.
  • Secured Ethernet-based network
    communication between all units.
  • Unique secured authentication of
    each unit to avoid fake/masking of alarms.
  • Remote viewing through a remote computer.
  • Gas-specific sensor with DISS nut and
    nipple to avoid mechanical cross-connection.
  • For Future Use- disables a gas alarm
    until the gas pipeline infrastructure
    is complete.
  • Multi-linguistic- Available in English,
    Spanish, Russian, Arabic & Chinese.
  • Commissioning- Enables to identify 
    gas leaks before the commission.
  • A color trend arc showing the gas
    pressure status.
  • Automatic adjustment of screen
    brightness (day & night mode).
  • Self-diagnostics with error display for
    problem identification.
  • Event logging, recording all alarms and
    panel activities.
  • Periodical maintenance alert.
  • Customizable text for panel location.
  • Highly accurate solid-state pressure
    piezo-resistive transducer.
  • Hinged panel opening for easy
    installation and maintenance.
  • On-site adjustment of HI and LOW
    set points for each Gas & gas unit.
  • Built-in relay for remote annunciator
    applications.
  • Gas-specific digital sensors mounted
    locally or remotely, up to 1500m,
    utilizing a 12 gauge twisted pair.
  • Wifi communication is available.
  • Remote monitoring (Via TCP/IP Modbus)
    through the building management system is available.

Applicable Standards

Medical Gas Systems:
NFPA99, 
EN ISO 7396-1 &
HTM02-01 
requirements.

  • The digital 12” touchscreen area
    alarm system shall be of Silbermann’s type.
  • The alarm Panel shall be capable of
    accommodating up to ten services.
    The sensors shall be capable of local
    or remote mounting.
  • Customizable language and gas badges
    (ISO/NFPA colors) during the setup process.
  • Each transducer shall be
    electronically gas-specific to avoid
    wiring cross-connection.
  • Each gas service shall have a unique
    display indicating the online measured
    pressure/vacuum.
  • Each gas service shall have a trend bar
    for showing the pressure trend.
    (Green - Normal, Yellow - Caution,
    Red - Hi or Low).
    In the case of an alarm, the red light
    will flash and the alarm will sound.
    Canceling the alarm by the mute button
    will stop the audible alarm but the red
    light will continue to flash until the failure is fixed.
  • TEST & SILENCE touch buttons
    shall be easily accessed to operate and test the unit.
  • Adjustable repeat alarm
    (15 or OFF. Also 15 seconds for
    convenient maintenance).
  • Selected gas pressure units: PSI, kPa, or BAR.
  • Selected vacuum pressure units: kPa,
    mmHg cmHg.
  • Alarm buzzer level adjustable: 70,80,90 dB
  • Adjustable Hi and Low set points
    (+/-20% of normal pressure).
  • Input power shall be 100VAC to 230VAC,
    50÷60Hz
  • Optional backup battery.