Silbermann 12” Touchscreen Area Alarm Panel for Real-Time Local Gas Info

The Silbermann Medical Area Alarm panel displays digitally and graphically the local medical gas information On a colorful 12” LCD touch screen.

This Xanadu touchscreen model uses Hi-security technology that protects the information displayed and transmitted from malicious attacks.

Xanadu Medical Gas Alarm System can receive information from other remote sources, alarm panels and manifold systems, and display it on the screen.

Effortlessly transmits all displayed information to any authorized device or system, maintaining secure and reliable communication across your facility

The Area Alarm panel complies with the NFPA99, EN ISO 7396-1, and HTM02-01.

Request a quote

Features and benefits

  • LCD 12” touch screen display.
  • Display of 1-10 gas alarms on screen.
  • Aesthetics & smooth design without buttons for infection control.
  • Secured Ethernet-based network communication between all units.
  • Unique secured authentication of each unit to avoid fake/masking of alarms.
  • Remote viewing through a remote computer.
  • Gas-specific sensor with DISS nut and nipple to avoid a mechanical cross-connection.
  • For Future Use- disables a gas alarm until the gas pipeline infrastructure is complete.
  • Multi-linguistic- Available in English, Spanish, Russian, Arabic & Chinese.
  • Commissioning- Enables to identify gas leaks before the commission.
  • A color trend arc showing the gas pressure status.
  • Automatic adjustment of screen brightness (day & night mode).
  • Self-diagnostics with error display for problem identification.
  • Event logging, recording all alarms and panel activities.
  • Periodical maintenance alert.
  • Customizable text for panel location.
  • Highly accurate solid-state pressure piezo-resistive transducer.
  • Hinged panel opening for easy installation and maintenance.
  • On-site adjustment of HI and LOW set points for each Gas & gas unit.
  • Built-in relay for remote annunciator applications.
  • Gas-specific digital sensors mounted locally or remotely, up to 1500m, utilizing a 12 gauge twisted pair.
  • Wifi communication is available.
  • Remote monitoring (Via TCP/IP Modbus) through the building management system is available.

 

Applicable Standards

Medical Gas Systems: NFPA99, EN ISO 7396-1 & HTM02-01 requirements.

Specifications

  • The digital 12” touchscreen area alarm system shall be of Silbermann’s type.
  • The alarm Panel shall be capable of accommodating up to ten services. The sensors shall be capable of local or remote mounting.
  • Customizable language and gas badges (ISO/NFPA colors) during the setup process.
  • Each transducer shall be electronically gas-specific to avoid wiring cross-connection.
  • Each gas service shall have a unique display indicating the online measured pressure/vacuum.
  • Each gas service shall have a trend bar for showing the pressure trend. (Green – Normal, Yellow – Caution,
    Red – Hi or Low). In the case of an alarm, the red light will flash and the alarm will sound. Canceling the alarm by the mute button will stop the audible alarm but the red light will continue to flash until the failure is fixed.
  • TEST & SILENCE touch buttons shall be easily accessed to operate and test the unit.
  • Adjustable repeat alarm (15 or OFF. Also 15 seconds for convenient maintenance).
  • Selected gas pressure units: PSI, kPa, or BAR.
  • Selected vacuum pressure units: kPa, mmHg cmHg.
  • Alarm buzzer level adjustable: 70,80,90 dB
  • Adjustable Hi and Low set points (+/-20% of normal pressure).
  • Input power shall be 100VAC to 230VAC, 50÷60Hz
  • Optional backup battery.

FAQ

What is a medical gas area alarm panel?
What are the benefits and importance of the panel size?
What gases are typically monitored?
Where are Area Alarm Panels typically installed
Are Area Alarm Panels customizable?
What standards do they comply with?
Can area alarm panels integrate with hospital management systems?
Is there an option for SMS alerts from an area alarm panel
Why is an area alarm panel important?
Is it possible to integrate analog gauges into area alarm panels?

Contact Us

    We'd love to hear from you!

    #5 Harakevet st. POB 4605, Petah-Tikva ISRAEL
    +972-3-9309186 info@silbermnn.com